Rule of 6s for Dialysis Access Placement

The "Rule of 6s" describes an easy way to evaluate the maturity of a recently-placed arteriovenous fistula for dialysis access--and is even mentioned by name in the most recent KDOQI Guidelines for Dialysis Access.  The Rule of 6s is as follows:

6 weeks after the AV fistula has been placed, the fistula should:
(a) be able to support a blood flow of 600 ml/min.
(b) be at a maximum of 6mm from the surface.
(c) have a diameter greater than 6mm.

Failure to achieve these goals warrants a further investigation, usually in collaboration with the access surgeon who placed the fistula, into why the fistula did not mature.  Ideally, the existing fistula can still be encouraged or modified to eventually achieve maturity; if not, a new access site can be attempted.  

4 comments:

  1. I think it has to be no more than 6 cm from the surface.

    ReplyDelete
  2. Chaisy,

    Quoting the K/DOQI guidelines

    "a depth of approximately 0.6 cm (ideally, between 0.5 to 1.0 cm from the skin surface). This combination of characteristics can be remembered easily as the Rule of 6s."

    0.6 cm = 6 mm

    6 cm would be much to deep to cannulate

    http://www.kidney.org/PROFESSIONALS/kdoqi/guideline_upHD_PD_VA/va_guide1.htm

    Sparks

    ReplyDelete
  3. Sorry I just misunderstood the whole thing. Thanks for clarifying..

    ReplyDelete
  4. There should be at least 6 inches vein avaialble for neelde puncture

    ReplyDelete

Renal Fellow Network encourages comments and discussion regarding the posts. Do not post any comments that are commercial or advertising in nature. Posts will be deleted if commercial or advertising comments are made. Internet users commenting on the Renal Fellow Network must post information which is true and correct to their knowledge. Sources to health/medical claims must be provided when relevant. Moderators reserve the right to erase, without notification, any comment they would judge inappropriate.