Poor phosphate control, or is it?


One of the commonest calls that one gets as a renal fellow in the dialysis unit concerns poor flows from catheters. This is often related to catheter-related thrombus formation and there are a number of different strategies employed to treat this complication and to prevent recurrence. In recent years, the use of alteplase to improve blood flows has become commonplace. In some patients who need repeated doses of alteplase to maintain catheter patency, locking the catheters with alteplase after dialysis has been shown to be effective in preventing thrombosis. Recently, however, an interesting side-effect of alteplase use in this setting has been reported.

An 83-year-old woman with ESRD was having occasional trouble with her dialysis catheter and had intermittently required alteplase-locking. She had been on a stable dose of PO4 binders and her PO4 was very well controlled. However, at one of her visits, her PO4 was noted to be 10.1 (previously 3.5-5) with no change in her diet and the following month it was 35.7mg/dl. Her labs were taken in the usual fashion, after discarding this first 5mls from the catheter. It was postulated that the alteplase might be interfering with the assay for PO4.

In order to confirm this, the authors withdrew 4 samples in 2ml increments from a catheter that had been locked with alteplase during the previous dialysis session. The PO4 concentration in the 4 samples was 161.5, 9.56, 6.47 and 5.68 mg/dl respectively. They then reviewed the records of a patient who had intermittent problems with hyperphosphatemia over the previous year. His mean PO4 level on days on which the catheter was not locked with alteplase was 6mg/dl compared to 9mg/dl on days with alteplase. In retrospect, his PO4 binders were increased inappropriately at least once in response to a spurious test.



So why did this happen? It turns out that the makers of alteplase use phosphoric acid for pH adjustment during the manufacturing process. Although the quantity is very small, it is enough to contaminate a draw from the catheter if it is not adequately removed beforehand. Spurious hyperphosphatemia has also been described in samples contaminated with heparin.

3 comments:

  1. The phosphorus levels can be quite dramatic. We have seen this in several of our pediatric patients who use catheters more regularly than adults.

    ReplyDelete
  2. One more reason why I draw my labs from the dialysis line just AFTER the treatment is initiated, not directly from the catheter itself. Too frequently the sample is contaminated with heparin, alteplase, or locking solution of choice, despite drawing wastage.

    ReplyDelete
  3. I agree, have seen falsely high Phosphates too, we sample just after blood has filled circuit. Also, TPA contamination can show falsely low PTH.

    ReplyDelete

Renal Fellow Network encourages comments and discussion regarding the posts. Do not post any comments that are commercial or advertising in nature. Posts will be deleted if commercial or advertising comments are made. Internet users commenting on the Renal Fellow Network must post information which is true and correct to their knowledge. Sources to health/medical claims must be provided when relevant. Moderators reserve the right to erase, without notification, any comment they would judge inappropriate.